CALABRATORS AND STANDARDS

Leak Detector Calibrators

calibrated leaksThe FCO99 series of calibrated leaks are designed to connect to production line pressure decay and low leak detectors such as the Furness Controls range of Leak Detectors. The calibrated leak device uses compressed copper material to present a leak at a certain test pressure. As there are no moving parts the device remains constant at a set pressure provided it is not contaminated by oil, dirt or water.

Models
FCO99-1 20cc/hr @150mBar
FCO99-2 1cc/min @1 Bar
FCO99-3 1cc/min @2 Bar
FCO99-4 2cc/min @4 Bar
FCO99-5 3cc/min @6 Bar
FCO99-6 5cc/min @8 Bar
FCO99-7 6cc/min @10 Bar
FCO99-8 10cc/min @12 Bar
FCO99-9 10cc/min @14 Bar

Digital MicrocalibratorThe FCO210 Digital Microcalibrator is a traceable flow rate meter designed primarily to provide calibration for production line leak detectors.
The unit comprises a mini flow meter using a laminar flow element and low range differential pressure transducer to resolve leak rate readings in cc/min or cc/hr. the leak rates are adjustable using the built in fine needle valve and the set leak is read on an LCD.

The FCO210 Microcalibrator is essential for the maintenance of leak detectors to optimum performance and assists in the traceability necessary for ISO 9000.

  • Ranges 0-20.00 cc/min and 0-2000 cc/hr
  • Direct readout in cc/min
  • Instantaneous response
  • Fine adjustment needle valve
  • Isolation valve
  • Provides traceability to National Standards
  • Connects to Leak Detector test or calibration port
  • Temperature compensated for variable ambient conditions

___________________________________